New test to improve safety of bionic devices

SYDNEY: Physicists at the University of Sydney have developed a method of testing if bionic devices — including pacemakers, cochlear hearing implants and bionic eyes

— are leak-proof and proposed new standards to ensure the devices can’t be penetrated by water.

If bionic medical devices are penetrated by moisture from bodily fluids and tissues, they could fail, which would put a patient’s health in jeopardy. While large leaks are easily detected during manufacturing, tiny leaks can escape detection.

Currently, implantable biomedical devices are moisturetested via a helium leak test, where helium gas is inserted into the device and the most sensitive method for detecting gas, known as mass spectrometry technology, is used to see if any escapes. But scientists have now found a way to measure gaseous or liquid water leakage without assuming that moisture behaves like helium. The new method, developed by David McKenzie from the University of Sydney, with Wenwen Lei from the National Measurement Institute, uses a standard leak based on the use of a stable polymer together with the most sensitive method for detecting the water, mass spectrometry. The unknown leak is quickly and reliably assessed against the standard leak that is supplied with a calibration.

In a paper published in a scientific journal of the American Chemical Society, Professor Mc-Kenzie and Dr Lei compared water leakage with helium leakage in biomedical test structures. They found that water can leak at up to 10 times the rate of helium, indicating that some devices could leak at too high a rate even after they had been leak-tested with the helium test.